1.1 Introduction
Capnograph adopts sidestream method to measure the end-idal carbon dioxide, airway respiration rate and other parameters. It can detect pulmonary ventilation function and reflect the circulation and pulmonary blood flow as well as indirectly reflect the alveolar ventilation standard. Meanwhile, it can also detect the blood oxygen saturation and pulse.
The Capnograph is applicable for use in pre-ospital emergency transfer, emergency department, ICU, operating room, neonatology, respiratory department and other occasions. The application population is adult, pediatric and neonate patients.
1.2 Features
- End-tidal CO2 measurement function
- Respiration rate measurement function
- SpO2 measurement function
- Pulse rate measurement function
- Over-limit alarm function for EtCO2, Respiration rate, SpO2 and Pulse rate
- Low battery alarm function
1.3 Performance
End-tidal CO2 measurement :
- Measurement range: 0 ~150mmHg
- Resolution: 1mmHg
- Accuracy :
- 0~40 mmHg:±2 mmHg
- 41~70 mmHg:±5% of readings
- 71~100 mmHg:±8% of readings
- 101~150 mmHg:±10% of readings
Respiration rate measurement:
- Measurement range: 2rpm~120rbpm
- Resolution: 1rpm
- Accuracy: ±1rpm
SpO2 measurement :
- Measurement range: 30~100 %
- Resolution: 1%
- Accuracy :
- 70~100 %:2 %
- 0 ~ 69 %:undefinition
Pulse rate measurement :
- Measurement range: 25~250 bpm
- Resolution: 1bpm
- Accuracy: ±2 bpm or ±2%,whichever is greater
Safety characteristic :
- Device type: internally powered equipment
- Degree of protection against electric shock: type BF applied part
- Degree of protection against ingress of liquid: IPX1
1.4 Accessories
- User Manual
- Blood oxygen probe
- Sampling pipe
1.5 Physical Characteristic
- Dimension: 159*72*31 mm
- Weight: 270 g
- Working environment
- Temperature: 5℃~40℃
- Relative humidity: 30%~ 75%, no condensation
- Atmospheric pressure: 700hPa~1060hPa
- Power supply: DC 3.7V (rechargeable lithium battries)
- Input power: ≤1.5VA